Technical note: False low turbidity readings from optical probes during high suspended-sediment concentrations

Nicholas Voichick, David J. Topping, and Ronald E. Griffiths

Download

Interactive discussion

Status: closed
Status: closed
AC: Author comment | RC: Referee comment | SC: Short comment | EC: Editor comment
Printer-friendly Version - Printer-friendly version Supplement - Supplement

Peer-review completion

AR: Author's response | RR: Referee report | ED: Editor decision
ED: Publish subject to minor revisions (review by editor) (08 Jan 2018) by Bettina Schaefli
AR by Anna Wenzel on behalf of the Authors (11 Jan 2018)  Author's response    Manuscript
ED: Publish as is (24 Jan 2018) by Bettina Schaefli
Download
Short summary
This paper describes instances in the Grand Canyon study area and a laboratory experiment in which very high suspended-sediment concentrations result in incorrectly low turbidity recorded with a commonly used field instrument. If associated with the monitoring of a construction or dredging project, false low turbidity could result in regulators being unaware of environmental damage caused by the actually much higher turbidity.